FDA-Device2022-07-20Class II

Draeger Globe-Trotter and Globe-Trotter IC, Model numbers MS20508 and MS20509. neonatal transport system

Draeger Medical Systems, Inc.
Hazard

The firm has initiated this recall to replace component on devices distributed containing a component within the oxygen supply system that is not rated to the working pressure of the system.

Sold states
No US distribution. Foreign distribution to BY, CL, DE, ES, GB, HU, IN, IT, KW, MX, SA, SV, TW.
Affected count
38 OUS (no US distribution)
Manufactured in
3135 Quarry Rd, Telford, PA, United States
Products
Draeger Globe-Trotter and Globe-Trotter IC, Model numbers MS20508 and MS20509. neonatal transport system

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1365-2022

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief