Moderate riskFDA DeviceClass II

Medica capillary tubes recalled for falsely high potassium test results

Medica Corporation's capillary tubes used with EasyBloodGas and EasyStat devices are giving false potassium results—up to 20% higher than actual levels. This could lead doctors to misinterpret a patient's potassium levels and provide incorrect treatment.

From the FDA notice · Read the official notice ↗

What to do

  1. Stop using Medica capillary tubes (Model 7303, Lot 25XXX)
  2. Contact Medica Corporation or your supplier for instructions on device return or replacement
  3. Notify your laboratory or healthcare facility of this recall

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Medica Capillary Tubes; EasyBloodGas/EasyStat Capillary Tube Kit.
Sold in
Nationwide
Made for
All ages (0 months and older)

Details

Units affected
1,578 units
Illnesses reported
None reported
Manufactured in
5 Oak Park Dr, Bedford, MA, United States
Classification
Class II
Recall number
Z-1370-2026

In the FDA's words

Description
Medica Capillary Tubes; EasyBloodGas/EasyStat Capillary Tube Kit. A systematic positive bias of up to 20% was identified across the reportable potassium range, resulting in false elevation results of patient samples when using Medica s capillary tubes. Model No. 7303, UDI-DI: 00840095607575, Lot No. 25XXX. Distribution: International distribution to the countries of Russia, Bulgaria, Kazakhstan, Ukraine, Pakistan, Belarus. Affected quantity: 1,578 units
Hazard
A systematic positive bias of up to 20% was identified across the reportable potassium range, resulting in false elevation results of patient samples when using Medica s capillary tubes.