Medica capillary tubes recalled for falsely high potassium test results
Medica Corporation's capillary tubes used with EasyBloodGas and EasyStat devices are giving false potassium results—up to 20% higher than actual levels. This could lead doctors to misinterpret a patient's potassium levels and provide incorrect treatment.
From the FDA notice · Read the official notice ↗
What to do
- Stop using Medica capillary tubes (Model 7303, Lot 25XXX)
- Contact Medica Corporation or your supplier for instructions on device return or replacement
- Notify your laboratory or healthcare facility of this recall
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- Medica Capillary Tubes; EasyBloodGas/EasyStat Capillary Tube Kit.
- Sold in
- Nationwide
- Made for
- All ages (0 months and older)
Details
- Units affected
- 1,578 units
- Illnesses reported
- None reported
- Manufactured in
- 5 Oak Park Dr, Bedford, MA, United States
- Classification
- Class II
- Recall number
- Z-1370-2026
In the FDA's words
- Description
- Medica Capillary Tubes; EasyBloodGas/EasyStat Capillary Tube Kit. A systematic positive bias of up to 20% was identified across the reportable potassium range, resulting in false elevation results of patient samples when using Medica s capillary tubes. Model No. 7303, UDI-DI: 00840095607575, Lot No. 25XXX. Distribution: International distribution to the countries of Russia, Bulgaria, Kazakhstan, Ukraine, Pakistan, Belarus. Affected quantity: 1,578 units
- Hazard
- A systematic positive bias of up to 20% was identified across the reportable potassium range, resulting in false elevation results of patient samples when using Medica s capillary tubes.