FDA-Device2019-05-29Class II

Lipid Panel test strips (as a component of smart bundles), REF 2729, also private labeled under Henry Schein as REF 570-0412

Polymer Technology Systems, Inc.
Hazard

Some lots of test strips do not fit tightly into the optical block, and may fall out during analysis, which could lead to an incorrect result being di P832 Chemistry (Diagnostic) CHH 2 Lipid Panel test strips (as a component of smart bundles), REF 2729, also private labeled under Henry Schein as REF 570-0412

Sold states
Domestic distribution nationwide.
Affected count
67643 total
Manufactured in
7736 Zionsville Rd, N/A, Indianapolis, IN, United States
Products
Lipid Panel test strips (as a component of smart bundles), REF 2729, also private labeled under Henry Schein as REF 570-0412

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1372-2019

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief