FDA-Device2019-05-29Class II
CHOL+HDL+GLU test strips, REF 2412
Polymer Technology Systems, Inc.
Hazard
Some lots of test strips do not fit tightly into the optical block, and may fall out during analysis, which could lead to an incorrect result being displayed
Sold states
Domestic distribution nationwide.
Affected count
67643 total
Manufactured in
7736 Zionsville Rd, N/A, Indianapolis, IN, United States
Products
CHOL+HDL+GLU test strips, REF 2412
Official notice