FDA-Device2023-04-19Class II

Edward Lifesciences, REF 777F8, Swan-Ganz CCOmbo V, CCO/SvO2/CEDV/VIP, Thermodilution Catheter, Exterior Diameter 8F (2.7mm), Usable Length 110 cm, Recommended Guidewire size 0.018 (0.46mm), Minimum Introducer size 9F (3.0 mm)

Edwards Lifesciences, LLC
Hazard

There is a potential that the distal end of the catheter may be bent beyond its intended curvature and when connected to patient monitoring device, the monitor may display an error message.

Sold states
US: AK AL AR AZ CA CO CT DC DE FL GA HI IA ID IL IN KS KY LA MA MD ME MI MN MO MS MT NC ND NE NH NJ NM NV NY OH OK OR PA RI SC SD TN TX UT VA WA WI WV WY OUS: None
Affected count
59,094 units
Manufactured in
1 Edwards Way, N/A, Irvine, CA, United States
Products
Edward Lifesciences, REF 777F8, Swan-Ganz CCOmbo V, CCO/SvO2/CEDV/VIP, Thermodilution Catheter, Exterior Diameter 8F (2.7mm), Usable Length 110 cm, Recommended Guidewire size 0.018 (0.46mm), Minimum Introducer size 9F (3.0 mm)

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1379-2023