FDA-Device2017-03-08Class II
PRODISC Orthopedic Manual Surgical Instrument, strut used in spinal surgery. Part numbers 03.820.350 and 03.820.352. Used to hold the disc segment open and facilitate discectomy
Synthes (USA) Products LLC
Hazard
These items are incorrectly etched and labeled. The item etched and labeled as 10mm Strut 03.820.350 lot 5919005 is a 12 mm Strut. The item etched and labeled as 12 mm Strut 03.820.352 lot 5919006 is a 10 mm Strut.
Sold states
US distribution to TX and LA
Affected count
2 units
Manufactured in
1301 Goshen Pkwy, N/A, West Chester, PA, United States
Products
PRODISC Orthopedic Manual Surgical Instrument, strut used in spinal surgery. Part numbers 03.820.350 and 03.820.352. Used to hold the disc segment open and facilitate discectomy
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1383-2017Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief