Covidien Signia Small Diameter Curved Tip Intelligent Reload 30 mm Vascular/Thin 8 mm - Short Item Code:SIGSDS30CTVT

Covidien, LP
Hazard

Potential for the device safety interlock to fail. The safety interlock prevents an empty single-use reload from being erroneously fired a second time. If the safety interlock fails and the empty single-use reload is fired a second time, the stapler may cut tissue but no staples will deploy from the empty reload, potentially resulting in a failed anastomosis, bleeding, tissue trauma, and the potential need to convert to an open procedure.

Sold states
Worldwide distribution - US Nationwide distribution and the countries of Austria, Belgium, Chile, China, Denmark, Finland, France, Germany, Hong Kong, Ireland, Italy, Netherlands, Norway, Poland, Portugal, Spain, Sweden, Switzerland, United Kingdom.
Affected count
15298 units WW: 9204 US; 6094 OUS
Manufactured in
60 Middletown Ave, N/A, North Haven, CT, United States
Products
Covidien Signia Small Diameter Curved Tip Intelligent Reload 30 mm Vascular/Thin 8 mm - Short Item Code:SIGSDS30CTVT

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1385-2021