FDA-Device2022-07-20Class III

HAMILTON-H900 Humidifier, Models: 950001, 950004

Hamilton Medical AG
Hazard

When starting a humidifier, intended for respiratory gas conditioning during invasive and noninvasive mechanical ventilation, with an empty chamber, and water is not filled into the chamber, the low water level alarm will be delayed and medical staff are unaware that there is no water in the chamber.

Sold states
US: WI, FL, CA, NV, PA, TX, PA
Affected count
203
Manufactured in
Via Crusch 8, Bonaduz, Switzerland
Products
HAMILTON-H900 Humidifier, Models: 950001, 950004

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1385-2022