FDA-Device2025-03-26Class II
Access 2 Immunoassay Analyzer, Catalog Number 81600N
Hazard
Beckman Coulter identified that an issue for Access 2 Sample Motors (B48016) due to improper alignment, causing the motors to rub against the sample wheelbase plate and index sensor. The issue was traced to a worn tool used in production (at the supplier), affecting Access 2 sample motors. This could lead to a potential delay in reporting patient test results.
Sold states
Worldwide - US Nationwide distribution.
Affected count
873 units
Manufactured in
1000 Lake Hazeltine Dr, N/A, Chaska, MN, United States
Products
Access 2 Immunoassay Analyzer, Catalog Number 81600N
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1386-2025More Beckman Coulter, Inc. recalls
- FDA-Device2026-04-15Beckman Coulter MicroScan Neg MIC 3J antibiotic susceptibility panels recalled for manufacturing defects
- FDA-Device2026-04-15MicroScan Neg Multidrug Resistant MIC 1 REF C32368 UDI-DI code: 15099590720568 Intended Use- For use with MicroScan Dried Gram Negative MIC/Combo Panels and Dried Gram Negative Breakpoint Combo Panels. MicroScan panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of aerobic and facultatively anaerobic gram-negative bacilli.
- FDA-Device2026-04-01Beckman Coulter MicroScan Neg MIC 56 panels recalled for potential well contamination