FDA-Device2018-04-25Class II
GE Healthcare Lunar: a) DPX Duo, Model Number: LU41693 b) DPX Bravo, Model Number: LU41692 Provides an estimate of BMD (Bone Marrow Density) at the lumbar spine and proximal femur regions. This BMD value can then be compared to a reference population at the sole discretion of the physician.
GE Medical Systems Ultrasound & Primary Care Diagnostics, LL
Hazard
Under certain conditions, when using DICOM Worklist along with DICOM MPPS, a report for a bone density exam may be sent to PACS with the incorrect patient information in the DICOM header. The correct patient information will be listed on the DICOM report image; however, the report may appear under a different patients name in the PACS.
Sold states
Worldwide Distribution - USA (nationwide) including Puerto Rico and to the countries of : Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Egypt, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Korea, Lebanon, Malaysia, Malta, Mexico, Morocco, Norway, Poland, Portugal , Romania, Russia, Saudi Arabia, Spain, Switzerland, Taiwan, Tunisia, Turkey, United Arab, United Kingdom and Venezuela.
Affected count
26 units
Manufactured in
3030 Ohmeda Dr, Madison, WI, United States
Products
GE Healthcare Lunar: a) DPX Duo, Model Number: LU41693 b) DPX Bravo, Model Number: LU41692 Provides an estimate of BMD (Bone Marrow Density) at the lumbar spine and proximal femur regions. This BMD value can then be compared to a reference population at the sole discretion of the physician.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1393-2018Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief