FDA-Device2022-07-27Class III

NeuRx Diaphragm Pacing System (NeuRx DPS), Affected part numbers: Part # Part Name 20-0035 NeuRx DPS System Kit 20-0036 NeuRx Diaphragm Pacing System, Patient Kit, SCI 20-0028 NeuRx DPS Surgical Kit, HDE 20-0040 NeuRx Diaphragm Pacing System Spare Sterile Components 21-0001-99 Mapping Probe 21-0004-99 Indifferent Electrode 21-0007-99 Lead Tunneler Set 21-0009-99 Blue Cable Set 21-0010-99 Transducer Tubing 21-0017-99 PermaLoc Electrode 21-0023-99

Synapse Biomedical Inc
Hazard

Firm has revised labeling to consistently call out storage conditions and revised the instructions for use to clarify that the temperature range symbol is intended to indicate storage conditions (versus transport conditions or tolerance conditions for the package material).

Sold states
Domestic distribution to AL CA CO FL GA IL MA MI MN NC NJ NY OH OK OR PA SC TX WV. Foreign distribution pending.
Affected count
71
Manufactured in
300 Artino St, N/A, Oberlin, OH, United States
Products
NeuRx Diaphragm Pacing System (NeuRx DPS), Affected part numbers: Part # Part Name 20-0035 NeuRx DPS System Kit 20-0036 NeuRx Diaphragm Pacing System, Patient Kit, SCI 20-0028 NeuRx DPS Surgical Kit, HDE 20-0040 NeuRx Diaphragm Pacing System Spare Sterile Components 21-0001-99 Mapping Probe 21-0004-99 Indifferent Electrode 21-0007-99 Lead Tunneler Set 21-0009-99 Blue Cable Set 21-0010-99 Transducer Tubing 21-0017-99 PermaLoc Electrode 21-0023-99 Green Cable 21-0024-99 Orange Cable 21-0025-99 Yellow Cable 21-0026-99 White Cable 77-0050 Surgeon Instruction Manual 20-0045 NeuRx DPS System Kit, ALS HDE 20-0046 NeuRx Diaphragm Pacing System, Patient Kit, ALS Marketed under HDE H070003 as well as H100006

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1393-2022

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief