FDA-Device2013-06-05Class II

Integrity 3.0 Product Usage: The Agility multileaf collimator is indicated for use when additional flexibility is required in conforming the radiation beam to the anatomy to be exposed.

Elekta, Inc.
Hazard

Unexpected movement during beam delivery may not be detected if there are incorrect static tolerances for Linac Geometric Axes loaded in Integrity R3.0.

Sold states
USA Nationwide including the states of CA, FL, IL, MD, MA, NJ, NY, NC, ND, OH, OR, PA, TN, TX, VA, WA, DC and Puerto Rico
Affected count
31
Manufactured in
400 Perimeter Center Ter Ne, Ste 50, Atlanta, GA, United States
Products
Integrity 3.0 Product Usage: The Agility multileaf collimator is indicated for use when additional flexibility is required in conforming the radiation beam to the anatomy to be exposed.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1402-2013