FDA-Device2021-04-21Class II

Chesapeake Cervical-Ti 10mm Drill Guide Inserter- a multi-screw construct providing stability to the anterior column (cervical spine) Catalog number: 3608-90033

K2M, Inc
Hazard

Interference condition between the 10mm Drill Guide Inserter and the inner shaft that mates with the inserters. As a result, inner shafts do not mate flush with the collet inserters, and splaying of the collet. If implant falls into the spinal cord may cause a spinal cord injury or Incomplete seating of the inserter to the implant could result in misalignment of the screws with the inserter screw guide, potentially causing malposition of the screws into the adjacent vertebral bodies

Sold states
PA
Affected count
1 unit
Manufactured in
600 Hope Pkwy SE, N/A, Leesburg, VA, United States
Products
Chesapeake Cervical-Ti 10mm Drill Guide Inserter- a multi-screw construct providing stability to the anterior column (cervical spine) Catalog number: 3608-90033

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1410-2021