Moderate riskFDA DeviceClass II

Medline Convenience Kits chest/breast-LF recalled for potential sterility issues

Medline Industries has recalled certain Chest/Breast-LF Convenience Kits (Model DYNJ83950) because equipment used to sterilize and package them had calibration problems. While the kits went through sterilization cycles, the calibration issues may have affected whether they meet required sterility standards. No injuries have been reported.

From the FDA notice · Read the official notice ↗ · All Medline recalls

What to do

  1. Check if you have a Medline Convenience Kit Model DYNJ83950 with lot numbers 24DBT832, 24CBH626, 23LBR075, 23FBF248, or 23CBG610
  2. Stop using any kits matching these lot numbers
  3. Contact Medline Industries or your supplier for instructions on return or replacement

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Medline Convenience Kits: 1) CHEST/BREAST-LF, Model Number: DYNJ83950
Sold in
Nationwide
Made for
All ages (under 100 years)

This recall covers 42 products — see all of them.

Details

Units affected
150 units
Illnesses reported
None reported
Manufactured in
3 Lakes Dr, Northfield, IL, United States
Classification
Class II
Recall number
Z-1412-2026

In the FDA's words

Description
Medline Convenience Kits: 1) CHEST/BREAST-LF, Model Number: DYNJ83950 Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products. 1) DYNJ83950, UDI-DI: 10195327318925(each), 40195327318926(case), Lot Number: 24DBT832; 2) DYNJ83950, UDI-DI: 10195327318925(each), 40195327318926(case), Lot Number: 24CBH626; 3) DYNJ83950, UDI-DI: 10195327318925(each), 40195327318926(case), Lot Number: 23LBR075; 4) DYNJ83950, UDI-DI: 10195327318925(each), 40195327318926(case), Lot Number: 23FBF248; 5) DYNJ83950, UDI-DI: 10195327318925(each), 40195327318926(case), Lot Number: 23CBG610 Distribution: Worldwide distribution - US Nationwide. Affected quantity: 150 units
Hazard
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

42 products in this recall

Medline convenience kits recalled for sterility assurance concerns

Medline is recalling 42 convenience kits and procedure packs that were sterilized and packaged on equipment with calibration problems. The kits went through the validated sterilization and packaging cycles, but the calibration issues could affect how reliably they are sterile. No injuries have been reported.

The agency lists each product separately. Check every one you might have.

More from Medline

See all

Related topics

SterilityMedical devicesSold nationwideMore from FDA Device