FDA-Device2020-03-11Class II

AIA-360 Automated Immunoassay Analyzer, Product Code 019945

Tosoh Bioscience Inc
Hazard

A display screen software issue on the analyzer causes the display to freeze when display screen is touched at the same time as a command from the instrument firmware, causing the instrument to stop. As a result, the run is aborted and the results are not retrievable.

Sold states
Distributed to AZ, NE, NJ, NY, and WA. Foreign distribution to Cayman Islands, Costa Rica, and Uruguay
Affected count
12
Manufactured in
3600 Gantz Rd, N/A, Grove City, OH, United States
Products
AIA-360 Automated Immunoassay Analyzer, Product Code 019945

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1420-2020

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief