FDA-Device2022-07-27Class II

Randox Cholesterol- For the quantitative in vitro determination of Cholesterol in serum and plasma. Catalog Number: CH2O0

Randox Laboratories Ltd.
Hazard

Product fails to meet the performance claims quoted on the kit insert: Manual Procedure when calibrating using the standard provided in the kit, the change in absorbance is lower than expected generating internal quality control results high outside range; Cholesterol (CHOL) Instrument Specific Application using the recommended calibration material CAL2351, the product fails to meet its linearity performance claim quoted on the kit inserts. The linearity is reduced by up to -35% however Quality Control results will still fall within assigned ranges. This may lead to delayed results.

Sold states
Nationwide Distribution - MD, OR, PR, MT, WV
Affected count
15 kits
Manufactured in
55 Diamond Road, N/A, Crumlin (North), N/A, Ireland
Products
Randox Cholesterol- For the quantitative in vitro determination of Cholesterol in serum and plasma. Catalog Number: CH2O0

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1424-2022