FDA-Device2013-06-05Class II

Vermed, Inc Radiotransparent Defibrillation Electrodes Monophasic or biphasic compatible defibrillation/cardioversion/pacing/ECG Monitoring electrodes. Used in connection with a defibrillator to deliver an electrical current (shock) to the patient's heart during cardiac emergencies.

Heart Sync, Inc
Hazard

On 4/18/13 HEARTSYNC became aware of a recall from their wire/connector sub-assembly supplier which is used in the C100-Physio Electrodes. There were reports of equipment damage due to misalignment from tooling installed in August 2012. The result of this damage could render the device incapable of delivering the appropriate energy or shock to the patient.

Sold states
Worldwide Distribution - USA including AL, GA, MI, MN, NC, ND, OH, PA, SD, VA, VT, OH, TX, MA, and Internationally to Korea.
Affected count
17180 units for all part numbers
Manufactured in
5643 Plymouth Rd, N/A, Ann Arbor, MI, United States
Products
Vermed, Inc Radiotransparent Defibrillation Electrodes Monophasic or biphasic compatible defibrillation/cardioversion/pacing/ECG Monitoring electrodes. Used in connection with a defibrillator to deliver an electrical current (shock) to the patient's heart during cardiac emergencies.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1435-2013