FDA-Device2020-03-11Class II

bellavista 1000 ventilator, Catalog number 301.100.030. bellavista is an electronically controlled pneumatic ventilation system with an integrated air compressing system. bellavista uses room air and high-pressure oxygen. Air enters through a fresh gas intake port and is compressed together with the oxygen by the blower. Oxygen enters through a high pressure inlet. An electronic mixer valve provides for the operator-set concentration. Gas is supplied to the patient via the microprocessor c

Vyaire Medical
Hazard

The G6 bellavista 1000 US ventilators may experience intermittent failures: Lack of acoustic high priority alarm, presence of a 'no alarm' condition, or presence of non-responsive touch screen.

Sold states
US Nationwide distribution in the states of AL, CA, CO, FL, GA, IL, MD, MI, PA, TX, WA and Puerto Rico.
Affected count
116
Manufactured in
26125 N Riverwoods Blvd, N/A, Mettawa, IL, United States
Products
bellavista 1000 ventilator, Catalog number 301.100.030. bellavista is an electronically controlled pneumatic ventilation system with an integrated air compressing system. bellavista uses room air and high-pressure oxygen. Air enters through a fresh gas intake port and is compressed together with the oxygen by the blower. Oxygen enters through a high pressure inlet. An electronic mixer valve provides for the operator-set concentration. Gas is supplied to the patient via the microprocessor controlled inspiratory valve.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1439-2020