FDA-Device2014-04-16Class I
Draeger Oxylog 3000 plus ventilators, emergency and transport ventilator. Part number: 5704833
Draeger Medical Systems, Inc.
Hazard
The potentiometers (knobs) for setting the ventilation parameters may have contact interrupted and the device discontinues ventilation, gives an audible alarm and displays the error message Poti unplugged.
Sold states
USA Nationwide Distribution in the states of WI and NY.
Affected count
3
Manufactured in
3135 Quarry Rd, N/A, Telford, PA, United States
Products
Draeger Oxylog 3000 plus ventilators, emergency and transport ventilator. Part number: 5704833
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1440-2014Don't want to check this manually?