FDA-Device2014-04-23Class II

AXIOM Vertix MD Trauma systems radiographic X-ray

Siemens Medical Solutions USA, Inc
Hazard

There is a potential issue and possible hazard to patients when using the AXIOM Vertix MD Trauma systems. In rare cases, steel ropes inside the lift column of the system can be defective without triggering the safety lock, which can result in the U-arm dropping down unexpectedly during movement in vertical direction, potentially causing serious injury.

Sold states
US Distribution including MO and OH.
Affected count
2
Manufactured in
51 Valley Stream Pkwy, N/A, Malvern, PA, United States
Products
AXIOM Vertix MD Trauma systems radiographic X-ray

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1460-2014