FDA-Device2018-04-25Class II
Cook Vacuum Pump ( Cook Aspiration Unit) K-MAR-5200-US, GPN G49275. Intended for the aspiration of body fluids and cells, in particular oocyte aspiration.
Cook Inc.
Hazard
Pump was found to be constructed from crimp connectors that do not meet the fixation wiring requirements and can cause failure of the device to operate or electric shock or burn to the user.
Sold states
Worldwide Distribution -- USA, Argentina, Bolivia, Brazil, Canada, Colombia, Guatemala, Mexico, Panama, Peru, Uruguay, and Venezuela.
Affected count
2,267
Manufactured in
750 N Daniels Way, Bloomington, IN, United States
Products
Cook Vacuum Pump ( Cook Aspiration Unit) K-MAR-5200-US, GPN G49275. Intended for the aspiration of body fluids and cells, in particular oocyte aspiration.