FDA-Device2021-04-28Class II

Manifold rotator (Component): a. 250 2 GANG MANIFOLD ROTATOR, OFF HANDLE, Model Number MX1232MR b. 250 3 GANG MANIFOLD ROTATOR, OFF HANDLE, Model Number MX1233MR c. 550 2 GANG MANIFOLD ROTATOR, OPEN HANDLE, Model Number MX1322MR d. 550 3 GANG MANIFOLD ROTATOR, OFF HANDLE, Model Number MX1333MR. Stopcocks and manifolds are used to control the direction of IV fluid flow. component of extravascular blood-pressure Transducer.

Smiths Medical ASD Inc.
Hazard

Specific lots of Medex High Pressure Adaptors, Rotators, Manifolds, Stopcocks and Tubing and Medex LogiCal Kits may have been manufactured with insufficient or incomplete welds. These are components for various cardiac kits including pressure monitoring devices and interventional imaging devices.

Sold states
Distribution US nationwide including Puerto Rico, Austria, Canada, Chile, Ecuador, Hong Kong, Mexico, Slovenia, Brazil, South Africa, Colombia, and Japan.
Affected count
1380 units
Manufactured in
6000 Nathan Ln N, N/A, Minneapolis, MN, United States
Products
Manifold rotator (Component): a. 250 2 GANG MANIFOLD ROTATOR, OFF HANDLE, Model Number MX1232MR b. 250 3 GANG MANIFOLD ROTATOR, OFF HANDLE, Model Number MX1233MR c. 550 2 GANG MANIFOLD ROTATOR, OPEN HANDLE, Model Number MX1322MR d. 550 3 GANG MANIFOLD ROTATOR, OFF HANDLE, Model Number MX1333MR. Stopcocks and manifolds are used to control the direction of IV fluid flow. component of extravascular blood-pressure Transducer.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1476-2021

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief