FDA-Device2018-05-16Class I

Fabius GS Premium, anesthesia machine, catalog no. 8607000 Product Usage: The products are inhalation anesthesia machines for use in operating, induction and recovery rooms. They may be used with O2, N2O, and AIR supplied by a medical gas pipeline system or by externally mounted gas cylinders. Fabius series anesthesia machines are equipped with an electrically driven and electronically controlled ventilator and monitors for airway pressure (P), volume (V), and inspiratory oxygen concentratio

Draeger Medical, Inc.
Hazard

The automatic ventilation may fail if the position detection of the motor is disturbed. If the ventilator cover is damaged, manual ventilation may also fail.

Sold states
US Nationwide Distribution to accounts in 21 states: AZ, CA, CO, ID, IL, KS, MA, MI, MS, NC, NY, OH, OK, PA, RI, TN, TX, UT, VA, WA, WI; and PR.
Affected count
27
Manufactured in
3135 Quarry Rd, N/A, Telford, PA, United States
Products
Fabius GS Premium, anesthesia machine, catalog no. 8607000 Product Usage: The products are inhalation anesthesia machines for use in operating, induction and recovery rooms. They may be used with O2, N2O, and AIR supplied by a medical gas pipeline system or by externally mounted gas cylinders. Fabius series anesthesia machines are equipped with an electrically driven and electronically controlled ventilator and monitors for airway pressure (P), volume (V), and inspiratory oxygen concentration (FiO2).

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1482-2018

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief