FDA-Device2013-06-12Class II

Zimmer Air Dermatome II, REF 00-8851-001-01 handpiece, w/o hose, Rx, non-sterile grafting instruments intended to provide variable graft thickness and width capabilities.

Zimmer Surgical Inc
Hazard

This removal is in response to complaints alleging that the Zimmer Air Dermatomes and Zimmer Air Dermatome II products would either not operate or operate intermittently. Zimmer's investigation determined that the planetary gear teeth were broken. This may cause a delay or postponement of the surgery or result in a suboptimal graft retrieval with potential of requiring an additional donor site.

Sold states
Worldwide distribution: USA Nationwide including Puerto Rico; UAE, TAIWAN, SWEDEN, THAILAND, KOREA, SINGAPORE, HONG KONG CHINA, CHILE, CANADA, AUSTRALIA, JAPAN, ENGLAND, GERMANY, and EGYPT.
Affected count
13 units
Manufactured in
200 W Ohio Ave, N/A, Dover, OH, United States
Products
Zimmer Air Dermatome II, REF 00-8851-001-01 handpiece, w/o hose, Rx, non-sterile grafting instruments intended to provide variable graft thickness and width capabilities.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1485-2013