FDA-Device2022-08-17Class I

Cobalt Implantable Cardioverter Defibrillators with Cardiac Resynchronization Therapy (CRT-D): a. Cobalt XT HF CRT-D, Model Numbers: DTPA2D4, DTPA2D1; b. Cobalt XT HF Quad CRT-D, Model Numbers: DTPA2QQ, DTPA2Q1

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Hazard

There is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).

Sold states
Worldwide Distribution
Affected count
8429 units
Manufactured in
8200 Coral Sea St Ne, Mounds View, MN, United States
Products
Cobalt Implantable Cardioverter Defibrillators with Cardiac Resynchronization Therapy (CRT-D): a. Cobalt XT HF CRT-D, Model Numbers: DTPA2D4, DTPA2D1; b. Cobalt XT HF Quad CRT-D, Model Numbers: DTPA2QQ, DTPA2Q1

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1486-2022