FDA-Device2022-08-17Class I

Cobalt Implantable Cardioverter Defibrillators: a. Cobalt VR ICD, Model Numbers: DVPB3D1, DVPB3D4; b. Cobalt DR ICD, Model Numbers: DDPB3D1, DDPB3D4

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Hazard

There is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).

Sold states
Worldwide Distribution
Affected count
2866 units
Manufactured in
8200 Coral Sea St Ne, Mounds View, MN, United States
Products
Cobalt Implantable Cardioverter Defibrillators: a. Cobalt VR ICD, Model Numbers: DVPB3D1, DVPB3D4; b. Cobalt DR ICD, Model Numbers: DDPB3D1, DDPB3D4

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1487-2022