FDA-Device2020-03-25Class II
Cardio Vascular-Allura Centron, system code 722400
Philips North America, LLC
Hazard
A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.
Sold states
Nationwide domestic distribution. Foreign distribution worldwide.
Affected count
87
Manufactured in
3000 Minuteman Rd, Andover, MA, United States
Products
Cardio Vascular-Allura Centron, system code 722400
Official notice