FDA-Device2020-03-25Class II

Cardio Vascular-Allura Centron, system code 722400

Philips North America, LLC
Hazard

A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.

Sold states
Nationwide domestic distribution. Foreign distribution worldwide.
Affected count
87
Manufactured in
3000 Minuteman Rd, Andover, MA, United States
Products
Cardio Vascular-Allura Centron, system code 722400

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1494-2020

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief