FDA-Device2017-03-29Class II

Giraffe Exam Light The Exam Light illuminates body surfaces and facilitates patient observation during a medical examination. It increases ambient light to facilitate performance of procedures on the mattress field. It is designed to mount on the accessory railings of the Giraffe OmniBed, incubators and warmers except for Panda and Giraffe Warmers. It is not intended for use to support surgical field lighting.

Ohmeda Medical
Hazard

Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where the power cord connects to the wall outlet which could cause thermal injury.

Sold states
Worldwide Distribution - US (nationwide and Puerto Rico) Bahamas and Internationally to UK, Trinidad & Tobago, Taiwan, Spain, Panama, Netherlands Antilles, Mexico, Lebanon, Japan, Ireland, Honduras, El Salvador, Ecuador, Costa Rica, Colombia, Canada, Brazil, and Venezuela,
Affected count
5,993 (5,080 US; 913 OUS)
Manufactured in
8880 Gorman Rd, Laurel, MD, United States
Products
Giraffe Exam Light The Exam Light illuminates body surfaces and facilitates patient observation during a medical examination. It increases ambient light to facilitate performance of procedures on the mattress field. It is designed to mount on the accessory railings of the Giraffe OmniBed, incubators and warmers except for Panda and Giraffe Warmers. It is not intended for use to support surgical field lighting.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1501-2017