FDA-Device2024-04-17Class II

EndoVive 20Fr Push Standard PEG Kit w/ENFit (Box 2)-Indicated for enteral nutrition directly into the stomach in both pediatric and adult patients who are unable to consume nutrition by conventional means. UPN: M00509051

Boston Scientific Corporation
Hazard

Inability to advance the guidewire through the feeding tube during placement due to blocked lumen of the barb connector. May lead prolongation in the procedure to exchange the device for another leading to leading to a cascade of events which may include additional intervention to close the puncture site, bleeding, leakage of gastric contents, and aspiration

Sold states
Worldwide distribution - US Nationwide and the country of Japan.
Affected count
572 units
Manufactured in
100 Boston Scientific Way, N/A, Marlborough, MA, United States
Products
EndoVive 20Fr Push Standard PEG Kit w/ENFit (Box 2)-Indicated for enteral nutrition directly into the stomach in both pediatric and adult patients who are unable to consume nutrition by conventional means. UPN: M00509051

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1514-2024