FDA-Device2014-05-07Class II
The Osteo-Clage System consists of stainless steel cable/sleeves and compression plates. . 2.0 mm x 850mm SS Cable/Sleeve; Part Number 01-0020-S 2.0 mm SS Sleeve Only; Part Number 01-0022-S 8, 7 hole, SS compression plate; Part Number 02-2007-S 10, 9 hole, SS compression plate; Part Number 02-2009-S Product Usage: The Osteo-Clage Stainless Steel Bone Plate is an implantable straight, rigid stainless steel plate. It is used in conjunction with 4.5mm Cortical bone screws to provide com
Acumed LLC
Hazard
Manufacturing of these devices with a grade of stainless steel that is not within specifications.
Sold states
US Nationwide Distribution - in the states of CA, CO, FL, GA, LA, MI, MN, MO, OH, OR, PA, PR, TX, UT, VA, WA and WY. Product was distributed outside the US to: Australia, China, Figi, France, Great Britain, Italy, Japan, Korea, Maylasia, Sweden, South Africa, Spain.
Affected count
141 units in the US; 294 units outside the US.
Manufactured in
5885 NW Cornelius Pass Rd, Hillsboro, OR, United States
Products
The Osteo-Clage System consists of stainless steel cable/sleeves and compression plates. . 2.0 mm x 850mm SS Cable/Sleeve; Part Number 01-0020-S 2.0 mm SS Sleeve Only; Part Number 01-0022-S 8, 7 hole, SS compression plate; Part Number 02-2007-S 10, 9 hole, SS compression plate; Part Number 02-2009-S Product Usage: The Osteo-Clage Stainless Steel Bone Plate is an implantable straight, rigid stainless steel plate. It is used in conjunction with 4.5mm Cortical bone screws to provide compression across bone fractures. The Osteo-Clage Cerclage Wire Crimp sleeve is used in conjunction with the bone plate in cerclage fixation procedures
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1515-2014Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief