FDA-Device2026-03-18Class IIPROCESSING DEFECT

Waldemar Link Endo-Model Replacement Plateau recalled for potential screw bushing detachment

OTHERNationwide distributionMedical devices

What this means

Waldemar Link GmbH & Co. KG is recalling certain Endo-Model Replacement Plateau implants (Item Number 15-8521/09) because the bushing on the plateau screw may detach from the screw shaft due to a crack. This could affect the stability of the implant. Eight units have been identified with this issue outside the United States.

  • If you received this implant, contact your surgeon or healthcare provider immediately
  • Provide your surgeon with your implant serial or lot number (250728/0038–0045)
  • Ask your surgeon whether your implant is affected and what monitoring or follow-up is needed
The agency's original wording
Hazard

The potential for an implant with a plateau screw where the bushing could detach from the screw shaft due to a longitudinal fracture.

Intended for
Adults
Illnesses reported
None reported
Sold states
Worldwide - US Nationwide distribution in the state of NJ and the countries of Belarus, Canada, Switzerland , Colombia, Germany, Spain, United Kingdom, Hungary, India, Italy, Netherlands, Romania, Saudi Arabia.
Affected count
8 units (OUS only)
Manufactured in
Ost Str. 4-10, N/A, Norderstedt, N/A, Germany
Products
Endo-Model Replacement Plateau; Item Number: 15-8521/09;

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1518-2026

Want us to check the next one for you?

We read every CPSC, FDA, USDA, and NHTSA notice and email you a plain-English Sunday brief covering your household — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Free · One email a week · Unsubscribe anytime