FDA-Device2021-05-12Class II
CRT-P Quad Models (CRT-Ps): a. Percepta CRT-P MRI: W1TR01, W1TR04, W4TR01, W4TR04 b. Serena CRT-P MRI: W1TR02, W1TR05, W4TR02, W4TR05 c. Solara CRT-P MRI: W1TR03, W1TR06, W4TR03, W4TR06. implantable medical devices used to treat arrhythmias related to bradycardia, and cardiac resynchronization therapy (CRT)
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Hazard
There is a SmartSync software issue that results in an overestimation in the displayed longevity of these devices during an approximate 6-month window of time before the device triggers its Recommended Replacement Time (RRT).
Sold states
Worldwide
Affected count
45,411 devices
Manufactured in
8200 Coral Sea St Ne, N/A, Mounds View, MN, United States
Products
CRT-P Quad Models (CRT-Ps): a. Percepta CRT-P MRI: W1TR01, W1TR04, W4TR01, W4TR04 b. Serena CRT-P MRI: W1TR02, W1TR05, W4TR02, W4TR05 c. Solara CRT-P MRI: W1TR03, W1TR06, W4TR03, W4TR06. implantable medical devices used to treat arrhythmias related to bradycardia, and cardiac resynchronization therapy (CRT)
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1523-2021Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief