Moderate riskFDA DeviceClass II

Waldemar Link Endo-Model Replacement Plateau recalled for detachable bushing screw

Waldemar Link GmbH & Co. KG is recalling specific lots of the Endo-Model Replacement Plateau (Item 15-0027/16), a surgical implant used in joint replacement procedures. The bushing on the plateau screw may detach from the screw shaft due to a crack, which could affect the implant's stability during or after surgery.

From the FDA notice · Read the official notice ↗ · All Waldemar Link recalls

What to do

  1. If you received this implant, contact your surgeon or healthcare provider immediately to determine if your device is affected.
  2. Do not undergo any procedure using this implant if your lot number matches: 2550412, 2550413, or 2549228.
  3. Your surgeon can verify the lot number on your implant paperwork or device packaging.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Endo-Model Replacement Plateau; Item Number: 15-0027/16;
Sold in
Nationwide
Made for
Adults (18 years to 100 years)

This recall covers 13 products — see all of them.

Details

Units affected
3 units (OUS only)
Illnesses reported
None reported
Manufactured in
Ost Str. 4-10, N/A, Norderstedt, N/A, Germany
Classification
Class II
Recall number
Z-1523-2026

In the FDA's words

Description
Endo-Model Replacement Plateau; Item Number: 15-0027/16; The potential for an implant with a plateau screw where the bushing could detach from the screw shaft due to a longitudinal fracture. Item Number: 15-0027/16; UDI-DI: 04026575432134; Serial/Lot Number: 2550412, 2550413, 2549228; Distribution: Worldwide - US Nationwide distribution in the state of NJ and the countries of Belarus, Canada, Switzerland , Colombia, Germany, Spain, United Kingdom, Hungary, India, Italy, Netherlands, Romania, Saudi Arabia. Affected quantity: 3 units (OUS only)
Hazard
The potential for an implant with a plateau screw where the bushing could detach from the screw shaft due to a longitudinal fracture.

13 products in this recall

Waldemar Link Endo-Model implant parts recalled for fracture risk

Waldemar Link is recalling 13 Endo-Model implant components, including replacement plateaus and connection components with PE-Plateau. The bushing on these implants could detach from the screw shaft because of a longitudinal fracture. No injuries have been reported.

The agency lists each product separately. Check every one you might have.