FDA-Device2020-04-01Class II

Karl Storz NEURO-FIBERSCOPE, Model # 11282BN1, Flexible Pediatric Neuroscope

Karl Storz Endoscopy
Hazard

During a complaint investigation it was determined that an error was observed within the instruction manual included with some scopes. Specifically, The Quick Connector Kits (QCK) and tray information, specified for the STERIS System 1E (SS1E) sterilization method within the instruction manual is incorrect. If the user uses SS1E to sterilize the impacted scopes and referred to the manual concerning the QCK information, the sterility of reprocessed scope cannot be guaranteed.

Sold states
US: AL, AZ, CA, CO, CT, DE, FL, GA, IL, IN, MA, MD, MI, MN, MO, NC, NH, NJ, OH, OR, PA, SC, TN, TX, VA, WA, WI, WV, OUS: Canda
Affected count
73 scopes
Manufactured in
2151 E Grand Ave, N/A, El Segundo, CA, United States
Products
Karl Storz NEURO-FIBERSCOPE, Model # 11282BN1, Flexible Pediatric Neuroscope

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1530-2020