FDA-Device2012-06-20Class II
Philips Healthcare Practix Convenio Mobile X-ray system Worldwide Distribution -- Nationwide
Philips Healthcare Inc.
Hazard
When the system is switched off or when the tube arm is parked in the down position (X-ray generator switch off), a radiation pulse (50 kV, 2.5 pGy) is generated inadvertently. .When the system is switched off or when the tube arm is parked in the down position (X-ray generator switch off), a radiation pulse (50 kV, 2.5 pGy) is generated inadvertently. The failure happened when component supplie
Sold states
Nationwide Distribution
Affected count
11 - US
Manufactured in
3000 Minuteman Road, Andover, MA, United States
Products
Philips Healthcare Practix Convenio Mobile X-ray system Worldwide Distribution -- Nationwide