FDA-Device2012-06-20Class II

Philips Healthcare Practix Convenio Mobile X-ray system Worldwide Distribution -- Nationwide

Philips Healthcare Inc.
Hazard

When the system is switched off or when the tube arm is parked in the down position (X-ray generator switch off), a radiation pulse (50 kV, 2.5 pGy) is generated inadvertently. .When the system is switched off or when the tube arm is parked in the down position (X-ray generator switch off), a radiation pulse (50 kV, 2.5 pGy) is generated inadvertently. The failure happened when component supplie

Sold states
Nationwide Distribution
Affected count
11 - US
Manufactured in
3000 Minuteman Road, Andover, MA, United States
Products
Philips Healthcare Practix Convenio Mobile X-ray system Worldwide Distribution -- Nationwide

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1538-2012

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief