FDA-Device2021-05-19Class I

Alaris Pump Module, Model 8100, serviced/repaired by Infusion Pump Repair with an impacted bezel repair part

Infusion Pump Repair
Hazard

Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may result in free flow, over infusion, under infusion or interruption of infusion.

Sold states
California
Affected count
15
Manufactured in
18 Technology Dr Ste 133, Irvine, CA, United States
Products
Alaris Pump Module, Model 8100, serviced/repaired by Infusion Pump Repair with an impacted bezel repair part

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1544-2021