FDA-Device2013-06-26Class II

Superior Approach Resection Guide Assembly. Indicated for primary, fracture, or revision total shoulder replacement for the relief of pain.

Biomet, Inc.
Hazard

Biomet has initiated this action following an investigation which identified that the slot on the resection guide is offset in the wrong direction. When the slot on the resection guide is offset in the wrong direction, the Steinman pin that is intended to hold the guide into place may not come into contact with the bone.

Sold states
Worldwide Distribution-USA (nationwide) including the states of UT, TX, AL, KY, and WI, and the countries of Australia, The Netherlands, Korea, and Canada.
Affected count
17
Manufactured in
56 E Bell Dr, Warsaw, IN, United States
Products
Superior Approach Resection Guide Assembly. Indicated for primary, fracture, or revision total shoulder replacement for the relief of pain.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1548-2013