FDA-Device2014-05-14Class II

Panda¿ iRes Infant Warmers integrated with Nellcor SpO2. Controlled infrared heat to neonates who are unable to thermo-regulate based on their own physiology with integrated monitoring feature used for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by a SpO2 sensor).

GE Healthcare
Hazard

Medical device software may be defective and may cause user-set alarm limits to result in false positive or false negative Oxygen saturation and pulse rate alarm notifications to the caregiver.

Sold states
Worldwide Distribution.
Affected count
723 units
Manufactured in
8880 Gorman Rd, N/A, Laurel, MD, United States
Products
Panda¿ iRes Infant Warmers integrated with Nellcor SpO2. Controlled infrared heat to neonates who are unable to thermo-regulate based on their own physiology with integrated monitoring feature used for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by a SpO2 sensor).

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1548-2014