FDA-Device2018-05-16Class I
AirLife Resuscitation, Adult, with mask, oxygen reservoir bag, CO2 detector Product Usage: The AirLife Resuscitation devices are manual resuscitation devices that are used to provide positive pressure ventilation to patients who are not breathing or cannot adequately breathe on their own.
Vyaire Medical
Hazard
The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
Sold states
Worldwide Distribution - US (Nationwide) and in the countries of Canada, Puerto Rico, United Arab Emirates, Philippines, and Europe.
Affected count
24 units
Manufactured in
75 N Fairway Dr, N/A, Vernon Hills, IL, United States