FDA-Device2024-09-18Class II

Olympus Triangle Tip Electrosurgical Knives- to be used with Olympus endoscopes and electrosurgical units to cut tissue using high-frequency current within the upper digestive tract Model: KD-645L

Olympus Corporation of the Americas
Hazard

Due to increase in complaints for the KD-640L and KD-645L Triangle Tip Electrosurgical Knives breaking off during use -issuing reminder to users to utilize the electrosurgical knife in accordance with the Instructions for Use, which details critical specifications regarding electrosurgical unit compatibility and output due to reports of serious injury

Sold states
Nationwide
Affected count
2525 units
Manufactured in
3500 Corporate Pkwy, Center Valley, PA, United States
Products
Olympus Triangle Tip Electrosurgical Knives- to be used with Olympus endoscopes and electrosurgical units to cut tissue using high-frequency current within the upper digestive tract Model: KD-645L

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1548-2024