FDA-Device2015-05-06Class II
2008 Series Hemodialysis Machines: 2008Kathome - 190895 2008T GEN 2 BIBAG WITHOUT CDX; 190908 2008T HD SYSTEM WITH CDX, CANADA; 190713 2008T HEMODIALYSIS SYS., with CDX; 190766 2008T HEMODIALYSIS SYSTEM W/BIBAG; 190858 2008T HEMODIALYSIS SYSTEM WITHOUT CDX The Fresenius 2008K is indicated for acute and chronic dialysis therapy.
Fresenius Medical Care Holdings, Inc.
Hazard
The Acetate value for GranuFlo on the Select Concentrate screen should be 8mEq/l but it is displayed as 4mEq/l. The correct value is shown on the dialysate screen.
Sold states
Worldwide Distribution: US (Nationwide) and countries of: Canada and Mexico.
Affected count
133,751 in total (US: 130,337 and OUS: Canada 1,201 and Mexico 2,213)
Manufactured in
920 Winter St, Waltham, MA, United States
Products
2008 Series Hemodialysis Machines: 2008Kathome - 190895 2008T GEN 2 BIBAG WITHOUT CDX; 190908 2008T HD SYSTEM WITH CDX, CANADA; 190713 2008T HEMODIALYSIS SYS., with CDX; 190766 2008T HEMODIALYSIS SYSTEM W/BIBAG; 190858 2008T HEMODIALYSIS SYSTEM WITHOUT CDX The Fresenius 2008K is indicated for acute and chronic dialysis therapy.
Official notice