FDA-Device2015-05-06Class II

2008 Series Hemodialysis Machines: 2008T with the following product code and description: 190395 2008K at Home System - 1st Gen. -¿ Canada; 190828 2008K at Home System - 2nd Gen - U.S.; 190904 2008K at Home System w/ Bibag; 190828 2008KatHOME HEMODIALYSIS SYSTEM; 190395 2008KatHOME MACHINE, SHORT CAB, OLC/DP, HP The Fresenius 2008K is indicated for acute and chronic dialysis therapy.

Fresenius Medical Care Holdings, Inc.
Hazard

The Acetate value for GranuFlo on the Select Concentrate screen should be 8mEq/l but it is displayed as 4mEq/l. The correct value is shown on the dialysate screen.

Sold states
Worldwide Distribution: US (Nationwide) and countries of: Canada and Mexico.
Affected count
133,751 in total (US: 130,337 and OUS: Canada 1,201 and Mexico 2,213)
Manufactured in
920 Winter St, Waltham, MA, United States
Products
2008 Series Hemodialysis Machines: 2008T with the following product code and description: 190395 2008K at Home System - 1st Gen. -¿ Canada; 190828 2008K at Home System - 2nd Gen - U.S.; 190904 2008K at Home System w/ Bibag; 190828 2008KatHOME HEMODIALYSIS SYSTEM; 190395 2008KatHOME MACHINE, SHORT CAB, OLC/DP, HP The Fresenius 2008K is indicated for acute and chronic dialysis therapy.