FDA-Device2015-05-06Class II
2008 Series Hemodialysis Machines: 2008K2 with the following product code and description: 190633 2008K2 HEMO SYSTEM, OPTIONAL MODULE; 190618 2008K2 HEMODIALYSIS SYS OLC/DP, SPANISH; 190610 2008K2 HEMODIALYSIS SYS. OLC/DIASAFE PLS; 190630 2008K2 MACHINE, SHORT CAB, OLC/DP, HP; 190628 2008K2 OLC ONLY WITH HEPARIN PUMP The Fresenius 2008K is indicated for acute and chronic dialysis therapy.
Fresenius Medical Care Holdings, Inc.
Hazard
The Acetate value for GranuFlo on the Select Concentrate screen should be 8mEq/l but it is displayed as 4mEq/l. The correct value is shown on the dialysate screen.
Sold states
Worldwide Distribution: US (Nationwide) and countries of: Canada and Mexico.
Affected count
133,751 in total (US: 130,337 and OUS: Canada 1,201 and Mexico 2,213)
Manufactured in
920 Winter St, Waltham, MA, United States
Products
2008 Series Hemodialysis Machines: 2008K2 with the following product code and description: 190633 2008K2 HEMO SYSTEM, OPTIONAL MODULE; 190618 2008K2 HEMODIALYSIS SYS OLC/DP, SPANISH; 190610 2008K2 HEMODIALYSIS SYS. OLC/DIASAFE PLS; 190630 2008K2 MACHINE, SHORT CAB, OLC/DP, HP; 190628 2008K2 OLC ONLY WITH HEPARIN PUMP The Fresenius 2008K is indicated for acute and chronic dialysis therapy.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1552-2015Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief