FDA-Device2019-06-05Class II
Maquet Getinge-BEQ-TOP 41200 ECC 1/4 PUMP PAC Material: 701063149
Maquet Cardiovascular, LLC
Hazard
Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured
Sold states
Nationwide
Affected count
N/A
Manufactured in
45 Barbour Pond Dr, N/A, Wayne, NJ, United States
Products
Maquet Getinge-BEQ-TOP 41200 ECC 1/4 PUMP PAC Material: 701063149