FDA-Device2015-05-06Class II

Caviwave Pro Ultrasonic Cleaning System, Product Usage: To wash (remove tissue, blood and other contaminants) or dry laparoscopic and general surgical instruments prior to final disinfection and sterilization as required.

Steris Corporation
Hazard

Should the Caviwave Pro Ultrasonic Cleaning System experience a temporary interruption in electrical power during a processing cycle, the circulation pump may stop functioning. In addition, it is possible that the ultrasonic generator may stop working. The machine does not provide an alert notifying the user of this device malfunction.

Sold states
Worldwide Distribution - US Nationwide in the states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NH,NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI & WV and the countries of: Australia, Brazil, Canada, France, Mexico, Saudi Arabia, Spain, Taiwan & United Arab Emirates.
Affected count
491 units
Manufactured in
5960 Heisley Rd, Mentor, OH, United States
Products
Caviwave Pro Ultrasonic Cleaning System, Product Usage: To wash (remove tissue, blood and other contaminants) or dry laparoscopic and general surgical instruments prior to final disinfection and sterilization as required.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1559-2015