FDA-Device2022-08-31Class I

Maquet Critical Care AB/Getinge Servo-u Ventilator System, Part Number 6694800 Servo-n Ventilator System, Part Number 6688600 Servo-u MR ventilator: Part Number 6888800 Servo-air: Part Number 6882000

Getinge Usa Sales Inc
Hazard

Ventilators may generate a combination of alarms which may result in loss of communication, technical alarms being triggered, and loss of ventilation. If the device stops ventilating due to this issue, it may lead to hypoventilation and consequently desaturation, with patient outcomes including hypoxemia and hypoxic injury, which might result in circulatory failure.

Sold states
Worldwide distribution. US nationwide, Poland, Spain, France, Germany, Czech Republic, Italy, Belgium, Estonia, Greece, Finland, Portugal, Netherlands, Romania, Austria, Slovakia, Lithuania, Ireland, Latvia, Norway, Sweden, Malaysia, China, Singapore, Taiwan, Iceland, Slovenia, Hungary, Bulgaria, Denmark, Croatia, United Kingdom, Russia, Hong Kong, United Arab Emirates, Canada, Switzerland, Australia, New Zealand, Serbia, Kenya, South Africa, Palestine, India, Colombia, Macau, Argentina, Mexico, Japan, Brazil, Israel, South Korea, Turkey, Peru, Ukraine, Thailand, Polynesia(French), Morocco, Saudi Arabia, Iran, Martinique (French), Costa Rica, Kuwait, Algeria, Saint Helena, Syria, Chile, Qatar, Fiji, Lebanon, Sri Lanka, Jordan, Oman, Pakistan, Ecuador, Botswana, Mozambique, Netherlands, Antilles, Bolivia, Andorra, Aruba, Faroe Islands, Egypt, Libya, Indonesia, Bahrain, Uruguay, Guadeloupe (French), Puerto Rico, Bosnia-Herzegovina, Paraguay, Nepal, Nicaragua, Vietnam, Tunisia, El Salvador, Tanzania, Papua New Guinea, Philippines, Nigeria, Myanmar, Dominican Republic, Cuba, Russian Federation, Honduras, Zimbabwe, Moldova, Republic of Bangladesh, Maldives, Angola Mauritius, Sudan, Turkmenistan, Panama, Rwanda, Brunei Darussalam, and Georgia.
Affected count
37000 (11089 US)
Manufactured in
1 Geoffrey Way, Wayne, NJ, United States
Products
Maquet Critical Care AB/Getinge Servo-u Ventilator System, Part Number 6694800 Servo-n Ventilator System, Part Number 6688600 Servo-u MR ventilator: Part Number 6888800 Servo-air: Part Number 6882000

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1560-2022