FDA-Device2022-08-31Class I

NeuRx Diaphragm Pacing System (NeuRx DPS) External Pulse Generator (EPG), Part No. 23-0021.

Synapse Biomedical Inc
Hazard

Due to a defect identified on the EPG controller printed circuit board assembly, device performance may degrade over time. Depending on the failure mode, potential adverse events include pain, muscle fatigue and decreased breathing effectiveness, arrhythmia or cardiac arrest, or the device may stop unexpectedly.

Sold states
distribution to US states of AL, CO, IL, FL, GA, MD, MN, NC, NJ, OH, PA, and SC; Germany and Kuwait. ***Updated 3/28/23 - Product was not distributed to United Arab Emirates. Product was distributed to Kuwait.*** Updated 9.14.23 - 1 patient in Germany, 1 patient in the Czech Republic, 1 distributor in Kuwait
Affected count
118 (112 US, 6 OUS)
Manufactured in
300 Artino St, Oberlin, OH, United States
Products
NeuRx Diaphragm Pacing System (NeuRx DPS) External Pulse Generator (EPG), Part No. 23-0021.