FDA-Device2020-04-01Class II

Captus 4000e Thyroid Uptake System, Model Numbers 5430-30151 and 5430-30152 - Product Usage: is intended to be used by trained nuclear Medicine Technologists or Nuclear Medicine Physicians to perform Thyroid Uptake Procedures.

Capintec Inc
Hazard

It is possible that the collimators have screw hole locations that were insufficiently tapped. There is a potential for the collimator assembly to fall and come into contact with a patient or operator.

Sold states
Worldwide distribution. US Nationwide, Australia, Austria, Bangladesh, Brazil, Canada, Chile, Colombia, Costa Rica, Ecuador, France, Hong Kong, India, Israel, Italy, Korea, Malaysia, Pakistan, Philippines, Poland, Saudi Arabia, Spain, Sweden, Taiwan, Thailand, Turkey, UAE, UK, and Vietnam. (list updated 3/26/2020)
Affected count
48 (481 additional units added 3/26/20)
Manufactured in
7 Vreeland Rd, N/A, Florham Park, NJ, United States
Products
Captus 4000e Thyroid Uptake System, Model Numbers 5430-30151 and 5430-30152 - Product Usage: is intended to be used by trained nuclear Medicine Technologists or Nuclear Medicine Physicians to perform Thyroid Uptake Procedures.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1564-2020