FDA-Device2014-05-14Class II

SpineFrontier S-Lift Lock Shaft Inserter, an instrument used with S-LIFT Intervertebral Body Fusion Device in spinal fusion surgery. Product Usage: The S-Lift Lock Shaft SI50001 mates with SI50002, Implant Inserter, to aid in securing the implant to the whole implant inserter assembly, SI50000.

SpineFrontier, Inc.
Hazard

It has been verified through field use that there are two possible breakages that may occur when using SI50001 Revision D Lock Shaft Inserter, possibly resulting in instrument fragments entering the sterile field: Impact Cap: The impact cap on the Lock Shaft Inserter can shear off under high-impact force during insertion and removal. Pin: The pin used to guide the Lock Shaft Inserter into the ma

Sold states
US Nationwide - OR, FL, MA, TX, CO, OK, IL, and MO
Affected count
30 devices
Manufactured in
500 Cummings Ctr Ste 3500, N/A, Beverly, MA, United States
Products
SpineFrontier S-Lift Lock Shaft Inserter, an instrument used with S-LIFT Intervertebral Body Fusion Device in spinal fusion surgery. Product Usage: The S-Lift Lock Shaft SI50001 mates with SI50002, Implant Inserter, to aid in securing the implant to the whole implant inserter assembly, SI50000.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1565-2014