FDA-Device2016-05-04Class II

Automated peritoneal dialysis (APD) cycler

Baxter Healthcare Corp.
Hazard

Baxter Healthcare Corporation is sending this communication to inform you of incomplete instructions in the AMIA Automated Peritoneal Dialysis (PO) System Clinician Guide for calculating the total recommended solution therapy volume. Specifically, the instructions do not specify the need for an extra 200 mL of PD solution in order to prime the patient line and for air purge operations. The tot

Sold states
US Consignee: NY ***Foreign Consignee: Ecuador
Affected count
10 units
Manufactured in
1 Baxter Pkwy, N/A, Deerfield, IL, United States
Products
Automated peritoneal dialysis (APD) cycler

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1565-2016

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief